Quantitative basis In photoluminescence, the emitted radiation is always dependent on the amount of radiation absorbed. Since an excited molecule returns to the ground state, it may generate energy loss in the form of no radiative transition. Therefore, the number of photons emitting radiation is usually less than the number of photons that absorb radiation. It is expressed by the quantum efficiency Q.
Under fixed experimental conditions, the quantum efficiency is a constant, usually Q is less than 1. For substances that can be used for fluorescent detection, the Q value is generally between 0.1 and 0.9. The fluorescence intensity F is proportional to the intensity of the absorbed light.
For dilute solutions, the fluorescence intensity is positively correlated with the concentration of the fluorescent substance solution, the molar absorptivity, the thickness of the absorption cell, the incident light intensity, the quantum efficiency of the fluorescence, and the fluorescence collection efficiency. The fluorescence intensity of a substance is directly proportional to the concentration of the substance under the condition that other factors remain unchanged, which is the quantitative basis of the fluorescence detector. Fluorescence detectors are solute detectors that can be used directly for quantitative analysis.
Excitation spectrum and emission spectrum fluorescence involve the two processes of light absorption and emission. Therefore, any fluorescent compound has two kinds of characteristic spectrum: excitation spectrum and emission spectrum.
Fluorescence belongs to photoluminescence, and it is necessary to select an appropriate excitation wavelength (Ex) to facilitate detection. The excitation wavelength can be determined by the excitation spectrum of the fluorescent compound. The specific detection method of the excitation spectrum is to scan the excitation monochromator so that the incident light of different wavelengths excites the fluorescent compound, and the generated fluorescence is detected by the light detection element through a fixed wavelength emission monochromator. The resulting fluorescence intensity versus excitation wavelength curve is the excitation spectrum. At the maximum wavelength of the excitation spectrum curve, the number of molecules in the excited state is the largest, that is, the absorbed light energy is also the highest, and the strongest fluorescence can be generated. When considering sensitivity, the determination should select the maximum excitation wavelength.
The so-called fluorescence spectrum actually refers only to the fluorescence emission spectrum. It is a curve that the fluorescence intensity of a monochromator for wavelength scanning is changed with the fluorescence wavelength (ie, the emission wavelength, Em) when the monochromator wavelength is fixed. Fluorescence spectroscopy can be used to identify fluorescent substances and serve as a basis for selecting an appropriate measurement wavelength for fluorescence measurement.
In addition, due to the characteristics of the fluorescence measuring instrument, the energy distribution of the light source, the transmittance of the monochromator, and the response of the detector will change with the wavelength, so that the same compound will have different spectra on different instruments, and No analogy with each other, this spectrum is called the apparent spectrum. To make the same compound have fluorescence spectra with the same characteristics on different instruments, the above characteristics of the instrument need to be corrected. The corrected spectrum is called a true fluorescence spectrum.
Excitation wavelength and emission wavelength are necessary parameters for fluorescence detection. Choosing the proper excitation wavelength and emission wavelength is very important for the sensitivity and selectivity of the detection, especially the detection sensitivity can be greatly improved.
Screw Chiller is called screw chiller because of its key component-compressor. The unit is made of gas refrigerant from the evaporator; after the compressor is adiabatically compressed, it becomes a high temperature and high pressure state. The compressed gas refrigerant is cooled and condensed in the condenser at equal pressure, and after condensing, it changes into a liquid refrigerant, and then expands to a low pressure through a throttle valve to become a gas-liquid mixture. The liquid refrigerant under low temperature and low pressure absorbs the heat of the material to be cooled in the evaporator and becomes gaseous refrigerant again. The gaseous refrigerant re-enters the compressor through the pipeline and starts a new cycle. The principle flow of screw chiller is divided into air cooling and water cooling.
Screw Chiller
Water Cooled Screw Chiller,Screw Water Chiller,Water Cooled Screw Compressor Chiller,screw top water cooler,screw water cooled liquid chiller,water cooled screw chiller trane,Water cooled flooded screw chiller, Screw type water chiller plant‎
Shenzhen Toyocool Refrigeration Equipment Co.,Ltd. , https://www.toyochiller.com